Appendix 8 Limitations Of Ultrasonic Monitoring; A8.1 How Does Ultrasound Work; A8.2 Artifacts In Fetal Heart Monitoring - EDAN F9 User Manual

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
Limitations of Ultrasonic Monitoring

Appendix 8 Limitations of Ultrasonic Monitoring

A8.1 How Does Ultrasound Work

When the ultrasound waves strike an object, they bounce back and create an echo. If the object
moves toward the sound source, the frequency of the echo increases. If the object moves away
from the sound source, the frequency of the echo decreases. This is called "Doppler Effect". In
the 1960's, the ultrasonic technique was first applied to medical diagnostic imaging.
The ultrasound process involves placing a small device called a transducer, against the skin of the
patient near the region of interest. The ultrasound transducer combines functions of emitting and
receiving ultrasounds in one device. This transducer produces a stream of inaudible, high
frequency sound waves which penetrate into the body and bounce off the organs inside. It detects
sound waves as they bounce off or echo back from the internal structures and contours of the
organs. The movement of the organs produces the Doppler Effect, and this movement can be
measured and described by measuring the echo.
In fetal monitoring, the ultrasound transducer produces a stream of sound waves which penetrate
into the maternal abdomen and bounce off the fetal heart. Then the transducer receives the echoes
and transfers them to the monitor, which turns the signal into fetal heart beating sound and fetal
heart rate trace.
Therefore, placement of the transducer is critical to ultrasound fetal heart monitoring.

A8.2 Artifacts in Fetal Heart Monitoring

(1) How does artifact happen?
The transducer detects sound waves as they bounce off or echo back from the fetal heart.
However, the sound waves bouncing off from maternal blood vessels may be detected by the
transducer and then be processed by the monitor as well. As a result, artifacts may be produced.
The artifacts, if not correctly interpreted, may cause the physicians to perform unnecessary
interventions, or to fail to detect the fetal distress and the need for interventions.
The most common artifacts are doubling and halving.
(2) Doubling:
When the FHR drops to 120 bpm or lower, the diastole and systole become far apart, thereby the
monitor may mistake these two movements of a single heartbeat for two separate heartbeats. As a
result, a heart rate trace that is double the actual heart rate is produced. This often happens during
severe decelerations and bradycardia, representing an abrupt switch of the trace to double the
actual heart rate.
- 201 -

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

F9 express

Table of Contents