EDAN F9 User Manual page 73

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
**TEMP LOW or
**TEMP xxx < yyy
Technical Alarm Messages
Alarm Message
High Level
***ECG SINGNAL
EXCEEDS LIMIT
Low Level
Signals
Overlap
(FHR1, HR)
Signals
Overlap
(FHR2, HR)
Signals
Overlap
(FHR1,
FHR2,
HR)
ECG LEADS OFF
ECG
EQUIP
MALF
HR
EXCEEDS
MEASUREMENT
RANGE
NIBP
EQUIP
MALF
NIBP
SYSTEM
FAILURE
NIBP
CUFF
LOOSE or OFF
TEMP result (xxx) is lower than
TEMP
the lower limit (yyy).
Source
Cause
ECG
signal
ECG
measurement limits.
US transducer 1 has picked up
US+ECG
the maternal heart signal; the
/Pulse
signals overlap.
US transducer 2 has picked up
US+ECG
the maternal heart signal; the
/Pulse
signals overlap.
US
transducer
US+ECG
transducer 2 have picked up the
/Pulse
maternal heart signal; the signals
overlap.
The ECG cable is not well
connected to the monitor or the
ECG
TOCO-E transducer, or the leads
of the ECG cable are not well
attached to the patient.
The
ECG
ECG
communicate with the system
successfully.
ECG/
The
heart
Pulse
measurement limits.
The
NIBP
NIBP
communicate with the system
successfully.
NIBP
The NIBP module defective.
The
cuff
NIBP
connected.
exceeds
the
1
and
US
board
cannot
rate
exceeds
the
board
cannot
is
loose
or
not
- 62 -
Countermeasure
Check the connection of the
leads
and
the
patient's
condition.
Reposition the US transducer
1 until the fetal heart signal is
detected.
Reposition the US transducer
2 until the fetal heart signal is
detected.
Reposition
the
transducers until the fetal
heart signals are detected.
Check
the
connection
between the ECG cable and
the monitor or the TOCO-E
transducer. Check also the
attachment of the ECG leads
to the patient.
Restart the monitor and try
again,
contact
manufacturer
if
connection still fails.
Check the connection of the
ECG leads/SpO
transducer
2
and the patient's condition.
Restart the monitor and try
again,
contact
manufacturer
if
connection still fails.
Restart the monitor and try
again,
contact
manufacturer
if
connection still fails.
Wrap the cuff properly.