F9, F9 Express Fetal & Maternal Monitor User Manual
Leakage Current
Ground
Leakage
(Limit)
Enclosure Leakage Current
(Limit)
Patient
Leakage
(Limit)
FHR1, FHR2, TOCO
Patient
Auxiliary
(Limit)
FHR1, FHR2, TOCO
Patient
Leakage
(Limit)
DECG, MHR(from MECG)
Patient
Auxiliary
(Limit)
DECG, MHR(from MECG)
Base Station
Weight
Size
US-T Transducer
Weight
Size
TOCO-T Transducer & TOCO-E Transducer
Weight
Size
N.C.
Current
500µA
N.C.
100µA
N.C.
Current
d.c. 10µA
a.c. 100µA
N.C.
Current
d.c. 10µA
a.c. 100µA
N.C.
Current
d.c. 10µA
a.c. 10µA
N.C.
Current
d.c. 10µA
a.c. 10µA
About1.8 kg
310mm x 235mm x81mm
About 150 g
81 mm × 35 mm
Ø
About 150 g
81 mm × 35 mm
Ø
S.F.C.
1000µA
S.F.C.
500µA
S.F.C.
50µA
500µA
S.F.C.
50µA
500µA
S.F.C.
50µA
50µA
S.F.C.
50µA
50µA
- 164 -
Product Specifications