Visual Alarm - EDAN F9 User Manual

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
is enabled again only when the SILENCE key is pressed.
If Alarm Reset is enabled (see 4.8 Pausing or Resetting the Alarm), and you press the SILENCE
key to disable an audible alarm, the alarm indicator will display
manner of lighting for one second and pausing for one second. When other alarms present, the
monitor will enable the audible alarm again automatically.
During the silence period, the alarm messages are displayed and the alarm indicator lights up as
usual. You can press the SILENCE key again to enable the audible alarm.
1
If the patient safety may be endangered, do not switch off the audible alarm infinitely.
2
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monitoring may result
in patient danger. Remember that the most reliable method of patient monitoring
combines close personal surveillance with correct operation of monitoring
equipment.
3
When the sound pressure of audible alarm is equivalent to the ambient noise, it may
be difficult for the operator to distinguish the audible alarm.
NOTE:
After you enable the audible alarm again, whether the alarm sound still exists depends
on whether the alarm persists.

4.3 Visual Alarm

When an alarm is active,
- the alarm indicator lights up:
Alarm Category
High level alarm
Medium level alarm
Low level alarm
- the alarm message appears in the message window of the main interface in yellow, with patient
alarms on the left and technical alarms in the middle.
- the numeric of the measurement flashes in grey with a frequency of 2Hz.
When more than one alarm of the same level is active, the alarm messages appear in the same
area in succession.
When more than one alarm of different levels are active, only the alarms of the highest level are
displayed in the message window.
The patient alarm messages are displayed either:
 in text form, for example "** FHR2 LOW"; or
 in numeric form, for example "** FHR2 115 < 120"; ** indicates this is a medium level
WARNING
Indicator Color
red
yellow
yellow
- 53 -
, which is flashing in the
Flashing Frequency
1.4Hz to 2.8Hz
0.4Hz to 0.8Hz
Constant (on)
Alarms
Duty Cycle
20% to 60% on
20% to 60% on
100% on