F9, F9 Express Fetal & Maternal Monitor User Manual
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/wireless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the F9 and F9 Express Fetal & Maternal Monitors are used exceeds the
applicable RF compliance level above, the F9 and F9 Express Fetal & Maternal Monitors should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the F9 and F9 Express Fetal & Maternal Monitors.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
c) The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur
radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4
MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz,21.0
MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
Table-Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
Test
Brand
Frequency
(MHz)
(MHz)
385
380-390
450
430-470
710
745
704-787
780
810
870
800-960
930
a)
a)
Service
TETRA 400
GMRS 460, FRS
460
LTE Brand 13, 17
GSM
800/900,TETRA
800, iDEN 820,
CDMA 850, LTE
a)
survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
equipment
Modulation
Maximum
b)
Power(W)
Pulse
b)
modulation
1.8
18Hz
C)
FM
± 5 kHz
deviation
1kHz sine
Pulse
b)
modulation
0.2
217 Hz
Pulse
b)
modulation
18 Hz
- 198 -
EMC Information
b)
Distance
(m)
0.3
2
0.3
0.3
2
0.3
IMMUNITY
TEST
LEVEL
(V/m)
27
28
9
28