Patient Id No. Setup - EDAN VE-H100B User Manual

Veterinary pulse oximeter
Table of Contents

Advertisement

VE-H100B Veterinary Pulse Oximeter User Manual
Set the PR limits as follows:
Choose PR Hi Alarm in the menu, press the Function
button to enter setup. The PR Hi Alarm box changes
from real line to broken line. The adjustable range of
the upper limit of PR Alarm is from "1 + the lower
limit of PR Alarm" to 350.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
Choose PR Lo Alarm in menu, press the Function
button enter setup. The PR Lo Alarm box changes
from real line to broken line. The adjustable range for
the lower limit of PR Alarm is from 0 to "the upper
limit of PR Alarm – 1".
Press the Function button, confirm the alarm range
setup.
Hi Alarm is always higher than Lo Alarm by at least
1 bpm.
Press the On/Off button to exit the menu, and return to
measurement state.

3.3.12 Patient ID No. Setup

The oximeter can support 100 IDs, and 300-hour data storage.
When entering menu, press the Function button to set ID (valid
range is from 1 to 100). The ID display box on the interface will
- 47 -