H100N Pulse Oximeter User Manual
on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
Choose SpO
Function button to set it. The SpO
change from real line box to broken line box. The
adjustable range for the lower limit of SpO
from 0 to "the upper limit of SpO2 Alarm - 1". If the
value of SpO2 Lo Alarm is set to less than 85, it will
restore to 85 after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
SpO
2
Alarm by at least 1%.
Press the Function button, and confirm the alarm
range setup.
Press the On/Off button to exit the menu, and
return to measurement state.
Lo Alarm in the menu, press the
2
Hi Alarm is always higher than SpO
Lo Alarm box will
2
- 59 -
Oximeter Operation
Alarm is
2
2
Lo