F9, F9 Express Fetal & Maternal Monitor User Manual
alarm event; the first number is the current measurement result; the second number is the
preset alarm limit.
The technical alarm messages are displayed in text form, for example "Fetus EQUIP MALF".
Setting alarm limits to extreme values may cause the alarm system to become
ineffective. It is recommended to use the default settings.
4.4 Choosing the Alarm Display Form
You can change the patient alarm display form,
1
Select the setup key
2
Select Alarm. On the displayed Password box, enter 9999, and then select Enter.
3
Select Message Form.
4
Select Text (default) or Numeric.
5
Select OK.
4.5 Changing the Alarm Volume
You can change the alarm volume,
1
Select the setup key
2
Select Alarm. On the displayed Password box, enter 9999, and then select Enter.
3
Select Alarm Volume.
4
Select Low (default), Medium or High.
5
Select OK.
4.6 *Choosing Alarm Silence Duration
You can change the alarm silence duration,
1
Select the setup key
2
Select Alarm. On the displayed Password box, enter 9999, and then select Enter.
3
Select Silence Duration.
4
Select Infinite, 1 min (default), 2 min, or 3 min.
5
Select OK.