EDAN F9 User Manual page 71

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
Medium Level
**Battery Low
Low Level
Check Paper
US1 UNPLUGGED or
US2 UNPLUGGED
US1 SIGNAL LOSS or
US2 SIGNAL LOSS
Fetus EQUIP MALF
TOCO UNPLUGGED
IUP UNPLUGGED
DECG LEADS OFF
DECG UNPLUGGED
DECG SIGNAL LOSS
DECG EQUIP MALF
Signals
Overlap
(FHR1, FHR2)
The battery power is too low to
Monitor
support further work of the
monitor.
There is no paper in the paper
Monitor
drawer or the drawer is open.
US
transducer
transducer
connected.
US
Or wireless US signal is not
detected.
FHR1 or FHR2 signal is too
US
weak
for
analyze.
The
fetus
US
communicate with the system
successfully.
TOCO transducer is not well
connected.
TOCO
Or wireless TOCO signal is not
detected.
IUP
cable
IUP
connected to the monitor.
The spiral electrode is not well
DECG
connected.
The DECG cable is not well
DECG
connected to the monitor or
the TOCO-E transducer.
DECG signal is too weak for
DECG
the system to analyze.
The
DECG
DECG
communicate with the system
successfully.
US transducer 1 and US
transducer 2 are aimed at the
US
same fetal heart; the signals
overlap.
- 60 -
Connect the monitor to AC
power supply.
Load paper and/ or close
the drawer.
1
or
US
2
is
not
well
Check the connection of the
transducer.
Check if the US transducer
is aimed at the fetal heart;
the
system
to
check if the alarm limits are
suitable; check the patient's
condition.
Restart the monitor and try
board
can
not
again,
manufacturer
connection still fails.
Check the connection of
both TOCO transducer and
US transducer.
is
not
well
Check the connection of the
IUP cable.
Check the connection of the
spiral electrode.
Check
between the DECG cable
and the monitor or the
TOCO-E transducer.
Check if the spiral electrode
is well attached to the fetus;
check
condition.
Restart the monitor and try
board
cannot
again,
manufacturer
connection still fails.
Adjust
transducers until another
fetal
detected.
Alarms
contact
the
if
the
the
connection
the
patient's
contact
the
if
the
one
of
the
US
heart
signal
is