F9, F9 Express Fetal & Maternal Monitor User Manual

7.5.3 Changing the UA Baseline

You can change the UA baseline,
1
Select the setup key
2
Select Fetus > UA Baseline.
3
Select 5, 10 (default), 15 or 20.
4
Select OK.
NOTE:
If the monitor has been configured with IUP, the IUP baseline is 0 and it is not adjustable.
The TOCO baseline is adjustable.

7.5.4 Testing TOCO Transducers

To test a TOCO transducer:
1
Switch on the monitor.
2
Connect the TOCO transducer to the fetal monitor.
3
Gently press the center of the transducer.
4
Check that the value on the display shows this change in pressure.
If a TOCO transducer fails the test, repeat this test with another transducer. If the second one
passes the test, defect of the first transducer is confirmed. Replace it with a good one. If the
second transducer fails the test as well, contact the manufacturer for service.
on the main interface.
Figure 7-7 Testing a TOCO Transducer
- 93 -
Fetal Monitoring