F9, F9 Express Fetal & Maternal Monitor User Manual
*
Non-linear Error
*
Resolution
Baseline
Temperature Changes
Zero Mode
Dielectric Strength
*
DFHR
Range
*
Resolution
*
Accuracy
*
Alarm
Technique
DECG
Input Impedance
Input Impedance
CMRR
Noise
Skin Voltage Tolerance
Fetal Input Voltage Current
*
Pressure Range
*
Non-linear Error
*
Resolution
IUP
Sensitivity
Zero Mode
*
Display Range
*
FM Mode
MFM
&
AFM
*
AFM Mode
AMF Technique
MHR Measurement Range 30bpm ~ 240bpm
*
MECG
±10%
1
1 unit/min/° C (free air)
Drift
due
to
5 units/min/° C (underwater)
Automatic (TOCO value becomes zero or below
lasting for 30 seconds)/ Manual
4000Vrms
30bpm ~ 240bpm
Measurement
1bpm
±1bpm
DFHR Alarm
Peak-peak detection technique
> 10M (Differential, DC50/60Hz)
> 20M (Common Mode)
> 110dB
< 4µVp
± 500mV
20µVp-3mVp
0mmHg ~100mmHg (0.0 kPa~13.3 kPa)
±3mmHg (±0.4 kPa)
1mmHg (0.1 kPa)
5µV/V/mmHg
Manual
0 ~ 999
Automatic/Manual
Trace (default) or Black Mark
Pulsed Doppler ultrasound
- 157 -
Product Specifications