System
Please wipe the transducer and the charging point with a dry cloth before charging the transducer.
Please do not scratch the charging point.
The battery is installed in the transducer. If the base station is switched on and supplied by AC,
the battery will be charged automatically when it is placed in the docking slot. Please make sure
the transducer is free of water and coupling gel before charging.
When you charge the battery, the screen will display as follows:
 Full charging icon: fully charged.
 Increasing charging icon: charging
 No charging icon: the transducer place in the docking slot incorrectly.
 If the screen displays ERROR, it indicates that the transducer is docked in the wrong base
station.
The transducer has a screensaver function:
When the transducer is fully charged in the docking slot for more than 10min (± 1min)
and the base station is supplied by AC, a small yellow full charging icon will be
displayed and floating on the screen and other icons will disappear.
When the transducer is charging in the docking slot for more than 10min (± 1min) but the
base station is not supplied by AC, a small yellow charging icon of real battery level will
be displayed and floating on the screen and other icons will disappear.
Introducing the FTS-3 Fetal Telemetry
- 46 -