Wireless Toco Monitoring - EDAN F9 User Manual

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
position should be different for different purposes: place the transducer close to the fetal
buttocks for NST, and place it on fetal back in delivery.
Place the transducer
on fetal buttocks for
NST
Wrap the abdomen with the belt over the transducer. Fix the transducer by pushing its buckle
through the overlapping section of the belt. Make sure the belt fits the patient snugly but
comfortably.
*Adjusting the Numeric to Zero
c)
Press the AUTO ZERO key to adjust the numeric to the baseline. Make sure this is not done
during a contraction.
Wipe off any gel presents on abdomen around this area.
NOTE:
1
Do not apply aquasonic coupling gel on a TOCO transducer or its contact area.
2
Check the function of the TOCO transducer by applying pressure on it to see if this
is displayed on the screen.

7.5.2 Wireless TOCO Monitoring

7.5.2.1 Parts Required
TOCO-T transducer or TOCO-E transducer
a)
Belt
b)
7.5.2.2 Monitoring Procedure
Connect the FTS-3 system to the monitor and switch on. Refer to section 3.2.7 connecting to
a)
the Monitor.
b) Take up the TOCO-T transducer or the TOCO-E transducer. Make sure it is successfully
connected to the base station.
c) Monitor TOCO following the procedures described in section 7.5.1.2 Monitoring Procedure.
Figure 7-6 Positioning TOCO Transducer
- 92 -
Fetal Monitoring
Place the transducer
on
fetal
back
in
delivery

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

F9 express

Table of Contents