Sterilizing - EDAN F9 User Manual

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
1 Do not use any disinfectant containing additional active ingredients other than those
listed.
2 Follow the manufacturer's instruction to dilute the solution, or adopt the lowest
possible density.
3 Do not immerse any part of the monitor or any accessory into liquid.
4 After disinfection, no remaining disinfectant is allowed on the surface.
5 Check if the monitor and accessories are in good condition. If any aging or damage is
detected (e.g. the belt loses its elasticity), replace the damage part(s) or contact the
manufacturer for service before reusing them.
6 Please do not light the TOCO transducer with
NOTE:
The manufacturer has no responsibility for the effectiveness of controlling infectious
disease using these chemical agents. Please contact infectious disease experts in your
hospital for details.
12.4

Sterilizing

Do not sterilize the monitor, the base station or the accessories, unless this is necessary according
to your hospital regulation.
NOTE:
Check if the monitor, base station, cables and accessories function well. If any problem
is detected, please contact the manufacturer for service before reusing them.
Checking Item
Checking Method
Visual
Inspect the monitor, base station and cables etc. for any damage.
Power On
Power on the monitor. Does it boot up successfully without errors and enter the
main menu?
Functionality Test
After power up, check whether the AC power indicator and battery status
indicator in the bottom left of the screen display as stated in section 2.4.1.
Performance
Please check the US transducer and TOCO transducer according to section
7.2.6 Testing US Transducers and section 7.5.4 Testing TOCO Transducers.
FTS-3 wireless transducers also can be tested accordingly.
System
When the monitor is connected to FTS-3, please check whether the base
station working channel and its battery status indicator in the bottom right of the
screen display as stated in section 2.4.1.
Type
NIBP Cuff
NIBP Cuff Extension Tube
Recommended
Ethanol 75%
Isopropanol 70%
CAUTION
ultraviolet light for a long time.