F15 Series Fetal & Maternal Monitor Service Manual

7.11 SpO2 Monitoring Failures

Phenomenon
No
SpO
waveform
2
numeric value.

7.12 NIBP Monitoring Failures

Phenomenon
The cuff fails to be inflated.
No NIBP numeric values.
Bad
connection
electrodes.
TOCO&MECG
defective.
Main
control
defective.
Possible Cause
SpO
transducer defective.
2
Bad connection of SpO
cable.
or
iM X12
parameter board
defective.
Main
control
defective.
Possible Cause
Bad connection of the cuff,
the air hose and the NIBP
socket.
Bad connection of the iM
X12 board to the NIBP
socket.
Cuff
or
extension
defective.
Air pump, air valve leak.
iM X12 parameter board
defective.
Main
control
defective.
iM X12 parameter board
defective.
Air pump, air valve leak.
Main
control
defective.
- 36 -
of
Check the connection.
Transducer
Replace
Transducer.
board
Replace the main control board.
Solution
Replace the SpO
2
Check the connection.
Replace the iM X12
.
board
board
Replace the main control board.
Solution
Check the connection from outside.
Check the connection from inside.
tube
Replace the damaged part(s).
Replace pump valve module.
Replace the iM X12 parameter board.
board
Replace the main control board.
Replace the iM X12 parameter board.
Replace pump valve module.
board
Replace the main control board.
Troubleshooting
the
TOCO&MECG
transducer.
2
parameter