F9, F9 Express Fetal & Maternal Monitor User Manual
completely out of the vagina, slide thumb between catheter and introducer tab, which
will begin to separate the introducer from the catheter.
9)
Anchor the catheter in place with one hand, and pull the introducer straight back off the
catheter.
10) Remove the liner from the adhesive pad, and then adhere the pad to the patient‟s skin.
Secure the catheter by placing the catheter attachment strap to the adhesive pad..
Rezeroing the System During Monitoring
1) With the catheter connected to the IUP cable, momentarily pressing the re-zero button
on the pressure cable. The green light on the cable will flash for five seconds.
2) During this period, adjust the monitor to zero by pressing AUTO ZERO key.
Figure 7-9 Separate the Introducer
Figure 7-10 Remove the Introducer
Figure 7-11 Secure the Adhesive Pad to Mother
Figure 7-12 Rezeroing the System
- 95 -