4.3.6 Volume Mode

4D Real Time 4D, RT_4D_Biopsy, VOCAL II, etc. are Options. If options are not part of the system
Note
configuration, checks can be omitted.
Different menus are displayed depending on which Touch Panel menu and which Volume Mode is selected.
Note
Some function keys only appear on the Touch Panel if they are available for the selected probe.
Note
4.3.6.1 Pre-Volume Mode Functions
1. Press the 3D or 4D key to select the desired Volume Mode.
The Volume Mode function is switched on, the "3D Pre" respectively "4D Pre" menu appears on the screen
(write mode) and the volume box appears on the Image area.
2. Select the Acquisition Mode (2).
3. Select an User Setting (3).
4. Select the desired Visualization Mode (4).
5. Start the Volume Acquisition with the Freeze key resp. the right trackball key.
Step
Task
1
ACQUISITION:
3D Static
4D Real Time
VCI A
VCI Omni View
STIC
Voluson E-Series Service Manual
KTD106657 Revision 2
Figure 4-16 Pre-Volume Mode menu (e.g., 4D Real Time)
Table 4-5 Pre-Volume Mode Functions
3D Volume Mode - Static volume acquisition (also in combination with PD, HD-Flow or
CFM)
Real Time 4D - continuous volume acquisition and parallel calculation of 3D rendered
images
Volume Contrast Imaging - improves the contrast resolution and the signal / noise ratio
and therefore facilitates finding of diffuse lesions in organs
Volume Contrast Imaging Omni View (any plane) - improves the contrast resolution and
the signal / noise ratio and therefore facilitates finding of diffuse lesions in organs.
Volumes can be edited in all other Visualization Modes.
The fetal heart or an artery can be visualized in 4D (also in combination with PD, HD-
Flow or CFM)
Functional Checks
Expected Results
4-17