Setup Instructions
3.7.1.5 How to activate Screen Lock
1.
Invoke System Setup as described in
2.
On the left side of the screen select
3.
Check mark "Screen Lock".
a.
If no password previously entered, following dialog appears.
b.
Enter "New Password".
c.
"Retype new Password" and then click
4.
Close the Service page with
A new screen lock password must be at least 6 characters long and has a maximum length of 80 characters.
Note
The password must contain at least 2 non-letter characters, 0...9 or !@#$%^*().
If screen is locked you have to enter the password to get full system control. If password is unknown click
Note
Emergency. This enables standard - but limited - operation.
Note
The Screen Lock password cannot be reset by the user! Please contact your GE service representative.
3.7.1.6 How to change Video Norm
1.
Invoke System Setup as described in
2.
On the left side of the screen select
3.
Click the proper field: PAL (50Hz) or NTSC (60Hz); see
4.
Close the Service page with
3.7.1.7 How to change Ext. Monitor Video Signal
1.
Invoke System Setup as described in
2.
On the left side of the screen select
3.
If not currently selected, click the proper field; see
4.
Close the Service page with
3.7.1.8 How to change Recorder Type
1.
Invoke System Setup as described in
2.
On the left side of the screen select
3.
Depending on the used Recorder type, click the proper field. Refer to
4.
Close the Service page with
3.7.1.9 How to adjust function of the Footswitch
1.
Invoke System Setup as described in
2.
On the left side of the screen select
3.
Select desired function of the Footswitch. Refer to
4.
Close the Service page with
3.7.1.10 How to change the Keyboard Layout
Section 6.4 "Modification of Keyboard Layout" on page 6-7
see
3-60
Section 3.7.1.1 on page 3-58
General and then click the General Settings tab.
Figure 3-61 change password
Save&Exit and restart the system.
Section 3.7.1.1 on page 3-58
Connectivity and then click the Peripherals tab.
Save&Exit and restart the system.
Section 3.7.1.1 on page 3-58
Connectivity and then click the Peripherals tab.
Save&Exit and restart the system.
Section 3.7.1.1 on page 3-58
Connectivity and then click the Peripherals tab.
Save&Exit and restart the system.
Section 3.7.1.1 on page 3-58
Connectivity and then click the Peripherals tab.
Save&Exit and restart the system.
.
Save&Exit to save new screen lock password.
.
Figure 3-49 on page 3-50
.
Figure 3-49 on page 3-50
.
Figure 3-49 on page 3-50
.
Figure 3-49 on page 3-50
.
.
.
.
Voluson E-Series Service Manual
KTD106657 Revision 2