Replacement Procedures
8.4 Image Settings Only - Loading Procedure
Introduction
The image settings contain:
•
Application Settings
•
2D Factory and 2D User Presets
•
3D/4D Factory and 3D/4D User Presets
•
Annotation Presets
•
Scan Assistant Configuration
•
Measure Configuration
•
Biopsy Lines
Loading Procedure
Section 4.4.2 "Load Small Backup (Scan Settings)" on page 4-31
see:
8.5 Full Backup (Full System Configuration) - Loading Procedure
Introduction
The Full Backup contains following data:
•
User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text entries,
gray curves and complete System settings such as language, time/date format, etc.)
•
Measure Configuration (user specific measure setup settings)
•
Patient Archive (database containing patient demographic data and measurements) - no images
•
Options (Permanent Key that is specific for enabled software options and Demo Key)
•
Image Transfer Configuration (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
•
Network Configuration (network settings including the computer name)
•
Service Platform (state of the Service Software)
Loading Procedure
Section 4.4.4 "Load Full System Configuration (Full Backup)" on page 4-36
see:
8.6 Image Archive - Loading Procedure
Introduction
A backup of the Image Archive contains the Patient Archive (database containing patient demographic data
and measurements) + images.
Loading Procedure
see:
Section 4.4.6.2 "Load Image Archive" on page 4-41
8-14
Voluson E-Series Service Manual
KTD106657 Revision 2