Replacement Procedures

8.4 Image Settings Only - Loading Procedure

Introduction
The image settings contain:
Application Settings
2D Factory and 2D User Presets
3D/4D Factory and 3D/4D User Presets
Annotation Presets
Scan Assistant Configuration
Measure Configuration
Biopsy Lines
Loading Procedure
Section 4.4.2 "Load Small Backup (Scan Settings)" on page 4-31
see:

8.5 Full Backup (Full System Configuration) - Loading Procedure

Introduction
The Full Backup contains following data:
User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text entries,
gray curves and complete System settings such as language, time/date format, etc.)
Measure Configuration (user specific measure setup settings)
Patient Archive (database containing patient demographic data and measurements) - no images
Options (Permanent Key that is specific for enabled software options and Demo Key)
Image Transfer Configuration (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
Network Configuration (network settings including the computer name)
Service Platform (state of the Service Software)
Loading Procedure
Section 4.4.4 "Load Full System Configuration (Full Backup)" on page 4-36
see:

8.6 Image Archive - Loading Procedure

Introduction
A backup of the Image Archive contains the Patient Archive (database containing patient demographic data
and measurements) + images.
Loading Procedure
see:
Section 4.4.6.2 "Load Image Archive" on page 4-41
8-14
Voluson E-Series Service Manual
KTD106657 Revision 2