8.17.2 Mounting/Replacing the VGA Image (Video) Resizer
The VGA Image (Video) Resizer is required whenever the used Secondary "Patient" Monitor has a different
Note
screen resolution than the Voluson E-Series system!
8.17.2.1 VGA Image Resizer - Installation Procedure
Note
Following steps describe, how to install the VGA Image Resizer (if it was not mounted before).
1.
Power Off/Shutdown the system as described in
2.
Open the right side door (A) of the Voluson E-Series system.
3.
Connect the power supply of the VGA Image Resizer to the power cable plug behind the door (1).
4.
Lead the cable through an opening (2) and close the door.
5.
Peel off the protective film from the adhesive tapes (on bottom of Image Resizer Box) and fix the image
resizer box on the Top Cover of the Voluson E-Series system (B).
6.
Connect the RGB Video Cable from the VGA Out connector (on Voluson E-Series system) to the
PC/HD IN connector on the Image Resizer box (3).
Place the cable in a way, that it can not get damaged when moving the Voluson E-Series system, or when
Note
the height of the User Interface gets adjusted.
7.
Connect the DC-Power Output from the Power Supply to the DC 12V Input connector at the Resizer
box (4).
8.
Connect your Secondary Monitor to the Image Resizer box at its PC/HD OUTPUT connector (5).
Figure 8-36 connector for Secondary Monitor
9.
Power On/Boot Up the system as described in
10.
Adjust the VGA Image Resizer settings as described in
11.
Measure Leakage Currents according to IEC 60601-1 respectively UL 60601-1.
Voluson E-Series Service Manual
KTD106657 Revision 2
Section 4.2.2 on page 4-4
Figure 8-34 power cable plug behind door
Figure 8-35 VGA video connection
Section 4.2.1 on page 4-3