Calibration Parameters page

Use this page to define the standards used for the calibration, to measure the
standards, and to select the reslope.
Standard conc. 0
This field is set to zero, and is not editable, but enables the blank to be measured.
Standards 1...5
Defines the concentrations of the calibration standards to be used. The value must
be greater than zero and have four significant figures, with up to 3 decimal places.
No two standards may have the same value, and they must be entered in
increasing order. Entering a value of zero will delete the current and subsequent
standards.
The reslope standard is set by selecting the desired standard and pressing the
Reslope key. The standard selected will be marked with an asterisk beside it to
indicate it is the Reslope standard. See 'Reslope the calibration' on Page 87 for
more information.
NOTE
If SIPS is enabled, the cursor is restricted to the last entry field, which is the
concentration of the bulk standard. Standards 1-4 are calculated from the diluted
bulk standard. The reslope standard is fixed at 3.

Results page

The Results page displays the solution results in tabular format, and displays
sample name, concentration, %RSD or precision, mean solution reading, dilution
factor and the last replicate value.
Figure 26. The Results page
64
Agilent 55B AA Spectrometer User's Guide