Maintenance
9
Figure 58. Impact adjustment block removed from the nebulizer block
Replace the impact bead
1
Figure 59. Align holes to enable assembly with the Impact bead adjustment screw
2
Figure 60. Insert the impact bead adjustment screw
104
Carefully push the broken portion of the Impact bead out from the Impact
bead adjustment block.
Insert the Impact bead adjustment block into the bottom of the nebulizer
block.
Ensure the hole in the Impact bead adjustment block is aligned with the
opening in the nebulizer block. This enables the assembly of the Impact bead
adjustment screw.
Replace the Impact bead adjustment screw by gently pushing past the O-ring.
Agilent 55B AA Spectrometer User's Guide