Operation
Figure 30. D
3

Align the burner

1
2
Figure 31. Squeeze prongs of rotation handle to rotate burner
76
lamp adjustment knobs
2
b
If the signal decreases, turn the knob in the other direction.
c
If the D
lamp signal is too small:
2
Check that the D
If the problem persists, press Enter to Rescale. This will bring the
signal back into range for display.
A signal that becomes too large should also be rescaled.
Repeat the previous step with the second D
Allow approximately 10-15 minutes for the hollow cathode lamp to warm up
and ensure nothing is in the optical path.
Rotate the burner by squeezing the prongs of the rotation handle, until the slot
is parallel to the light path.
lamp is installed correctly. (refer to Page 124).
2
2
Agilent 55B AA Spectrometer User's Guide
lamp adjustment knob.