Maintenance
NOTE
Do not force the white plastic insert. It may be necessary to screw
counter-clockwise first for lead-in thread to be located.
8
Figure 50. Insert the clamp bush guide into the Nebulizer block.
9
10 Check the spring is operational by pushing the capillary assembly with a screw
Figure 51. Check spring function
11 Thread the thimble and locking ring over the capillary.
12 Screw the thimble onto the nebulizer body.
Instructions on how to adjust the Nebulizer are detailed in 'Nebulizer setup' on
Page 81.
100
Place the clamp bush guide, thread end first, in the nebulizer body.
Screw the clamp bush guide into the white plastic insert using the small end of
the Nebulizer tool. Do not over-tighten.
driver. The assembly should move in and out freely.
Agilent 55B AA Spectrometer User's Guide