9337 Instrument Fault 310V Psu; 9339 Mono Resetting. Please Wait; 9415 Instrument Fault: Lamp Code; 9422 Instrument Fault: Lamp Current - Agilent Technologies 55B AA User Manual

Spectrometer
Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

A wavelength correction fault. The error 9621 Instrument Fault: Wl correction
occurred prior to this error.
Suggested action:
Check that the optical light path is not obstructed in the sample compartment.
Correct any of the possible causes listed above. Exit, then re-enter the Optimize
page/window. This will cause the instrument to retry the peaking.

9337 Instrument fault 310V PSU

Instrument Fault. The instrument detected a problem with the 310 volt supply on
the control board. A service call may be required.

9339 Mono resetting. Please wait

This message is for information only. It can only occur when the instrument is just
switched on. If this message appears then the instrument will take several
minutes to find its wavelength reference. The instrument will function correctly
after the delay. If this message continues to occur and the startup delay is a
problem, then a service call will be required.

9415 Instrument fault: lamp code

The instrument could not recognize the lamp code for one of the installed hollow
cathode lamps.
Possible causes include:

9422 Instrument fault: lamp current

The instrument could not detect any lamp current for the selected hollow cathode
lamp.
Possible causes include:
Agilent 55B AA Spectrometer User's Guide
Slit width (SBW) selection incorrect.
The lamp position in the method is incorrect.
One of the installed hollow cathode lamps is faulty.
A electronic fault in the instrument's lamp recognition circuity.
No hollow cathode lamp is installed in the selected lamp position.
Faulty hollow cathode lamp.
Faulty lamp supply circuits.
Troubleshooting/Errors