Agilent Technologies 55B AA User Manual page 127

Spectrometer
Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

4
5
6
NOTE
If a fuse blows repeatedly, it may indicate other problems with the
instrument. Contact your local Agilent sales office and arrange a service
call.
Agilent 55B AA Spectrometer User's Guide
Gently lever the fuse holder (2) from the instrument.
Check that the fuses are the correct type and are not damaged. Replace if
necessary.
Replace the fuse holder in the instrument and reconnect the instrument to the
mains power supply.
Maintenance
127