Agilent Technologies 55B AA User Manual page 105

Spectrometer
Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

3
4
Figure 61. Tighten the trapping screw – gas inlet side of the nebulizer
5
6
Figure 62. Tighten the clamping screw – side of nebulizer
7
This restricts the horizontal movement of the Impact bead adjustment block.
Figure 63. Clamp screw touching the impact bead adjustment block
Agilent 55B AA Spectrometer User's Guide
Turn the Impact bead adjustment screw until it engages with Impact bead
adjustment block.
Tighten the Trapping screw clockwise as far as possible. Do not over-tighten.
Loosen the Trapping screw one turn.
This enables free movement of the Impact bead adjustment screw.
Tighten the Clamping screw clockwise until it gently touches the Impact bead
adjustment block. Do not over-tighten.
Loosen the Clamping screw one-eighth of a turn.
Maintenance
105