9321 No peak: low HC lamp energy

Low hollow cathode lamp energy. Monochromator reference wavelength not
found.
The instrument Gain had been raised to its highest value but the hollow cathode
lamp signal was too low.
Possible causes include:
Suggested action:
Check that the optical light path is not obstructed in the sample compartment. Exit
then re-enter the Optimize page/window. This will cause the instrument to retry
the peaking.

9322 No peak: high HC lamp energy

High hollow cathode lamp energy. Monochromator reference wavelength not
found. The instrument Gain had been set to its lowest value but the hollow
cathode lamp signal was too high.
Possible causes include:

9323 Low emission: no peak

The instrument failed to peak at the selected wavelength. Low Emission energy at
the selected wavelength. The instrument Gain had been raised to its highest value
but the signal was too low.
Possible causes include:
Agilent 55B AA Spectrometer User's Guide
Hollow cathode lamp misaligned.
The light path is obstructed.
Faulty hollow cathode lamp.
Hollow cathode lamp misaligned.
Faulty hollow cathode lamp.
Faulty EHT circuit.
Flame not lit, or correctly aligned.
Wrong wavelength selected.
The standard solution currently being aspirated has low concentration.
Emission Setup has not been performed.
Faulty PMT or related circuits.
Troubleshooting/Errors