Maintenance
NOTE
When you fit a new capillary assembly, you should also fit a new capillary guide and
spring
When you fit a new venturi, replace all O-rings. Always replace O-rings as a set. If
you are using organics, ensure you fit the specific organic O-ring kit.
To reassemble the components of the nebulizer:
1
2
Figure 44. Insert Venturi into Nebulizer Block
Place the capillary guide into the nebulizer body, with the O-ring end up.
Figure 45. Insert Capillary Guide into the Nebulizer Block
3
4
98
Hold the nebulizer block front end up.
Place the venturi in the nebulizer body, thin end first.
Place the large end of the venturi extraction tool over the capillary guide.
Push down on the capillary guide and venturi.
The venturi should protrude from the opposite end of the nebulizer block.
Agilent 55B AA Spectrometer User's Guide