Agilent Technologies 55B AA User Manual page 102

Spectrometer
Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

Maintenance
Figure 53. Loosen the Impact locking screw - bottom of the nebulizer block
1. Impact bead locking screw
4
Figure 54. Glass Impact Bead removed from the back of the Nebulizer Block
NOTE
You should only perform the remainder of this procedure if the bead is broken and
trapped in the adjuster block.
5
102
2. Impact bead adjustment block
Using a slight twisting motion, remove the impact bead from the nebulizer
block.
Loosen the Trapping screw 3-4 turns.
The Trapping screw retains the Impact bead adjustment screw in the nebulizer
block.
3. Flat-blade screw driver
Agilent 55B AA Spectrometer User's Guide