Troubleshooting/Errors

9531 No lamp current detected

The instrument could not detect any lamp current for the selected Hollow Cathode
lamp and has aborted the peaking. Rectify the problem and optimize the
instrument.
Possible causes include:

9602 Instrument fault: mono datum

Instrument Fault. The instrument detected a problem with the monochromator.
Restart the instrument. If the problem persists a service call will be required.

9911 No burner fitted

An attempt was made to ignite the flame without any burner installed in the
sample compartment.
9912 N
Wrong burner fitted. Change to N
Possible causes include:
Wrong burner fitted. The burner fitted is an air-acetylene type, but an N
acetylene flame is requested.
Burner detection circuitry failure.

9914 No gas control unit

There is no flame control unit fitted to this instrument. An attempt was made to
load a flame method.

9915 Flame shield open

The flame shield was detected open or removed when the gas-box attempted to
ignite the flame.
150
No hollow cathode lamp in the selected lamp position.
Faulty hollow cathode lamp.
O burner not fitted
2
O-acetylene burner.
2