Agilent Technologies 55B AA User Manual page 59

Spectrometer
Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

NOTE
The element can only be selected by loading an appropriate method (user or
cookbook).
This page includes the following instrument-specific fields:
Instrument mode
Defines the mode you want to use, 'Absorbance' or 'Emission' - default is
Absorbance.
Active lamp
Defines which lamp position to use - (default is position 1).
Active current
Defines the current for the active lamp - default setting is the recommended
current for the lamp of the selected element.
Standby current
Defines the current for the standby lamp - default is 0.0.
D
correction
2
Defines whether D
Gas type
Defines the flame type to be used for the analysis - default is the recommended
gas type for the selected element.
Wavelength
Defines the wavelength to be used for the current element - default is the first
listed wavelength.
NOTE
The listed wavelengths are element and instrument-mode specific.
Other λ
Defines the wavelength to be used when 'Other' is selected in the Wavelength field.
Range is 180.0–900.0 nm.
Agilent 55B AA Spectrometer User's Guide
correction is to be used or not - default is 'No'.
2
Interface
59