Collect A Trouble Image With Logs - GE LOGIQ P6 Service Manual

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
7-2-3

Collect a Trouble Image with Logs

If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture
of the image monitor, system presets and the following logs:
-
-
-
-
-
NOTE:
Power Supply and Temperature logs are not currently being updated by the LOGIQ™ P6/P6
Pro.
This Alt-D function is available at all times.
When Alt-D is pressed, a menu box appears that allows for:
-
-
The subsequent file is compressed and time stamped. The screen capture is a bitmap which eliminates
the possibility of artifacts from compression.
3
Keyboard Shadow Log
Error Logs
Crash Log
Power Supply
Temperature
Figure 7-1 ALT-D Dialog Box
A place to enter a description of the problem
A choice to store to a pre-formatted CD-R/DVD, USB or Network storage.
Section 7-2 - Gathering Trouble Data
LOGIQ™ P6/P6 P
S
RO
ERVICE
M
ANUAL
7 - 3