Main Keyboard Assy
8-2-6-1
Tools
•
Common philips screwdrivers
8-2-6-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-6-3
Removal procedure
1.) Remove the 3 Probe holders & Cup holder from the keyboard. Refer to the Figure 8-70.
2.) Unscrew 4 screws from the bottom front of the keyboard. Refer to the Figure 8-71.
8 - 36
3
Figure 8-70 Removing holders
(1)
(2)
Figure 8-71 Unscrews 4 screws
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P6/P6 P
(4)
(3)
S