Common Diagnostics; Utilities - GE LOGIQ P6 Service Manual

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
Section 7-7

Common Diagnostics

Common Diagnostics is the first main heading under the GSUI heading of Diagnostics.
Figure 7-12 shows the top level menu choices under Common Diagnostics. The subheadings and their
respective diagnostic/troubleshooting menu choices are called out and described in the sections below.
7-7-1

Utilities

Provides two selections:
7-7-1-1
Disruptive Mode
Allows you to enable or disable disruptive mode troubleshooting. If you are accessing through InSite,
this can only be enabled with the customer/operator confirmation.
7-7-1-2
System Shutdown
Allows for system shutdown from the diagnostic menu. Select to retain Disruptive Mode or Not.
After submitting to or shutdown a confirmation screen gives one last chance to confirm or cancel the
request.
7 - 14
3
Figure 7-12 Common Diagnostics
Section 7-7 - Common Diagnostics
LOGIQ™ P6/P6 P
S
RO
ERVICE
M
ANUAL