GE LOGIQ P6 Service Manual page 266

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-13
Keycap set, Main Keyboard Encoder Knob set
8-2-13-1
Tools
Common philips screwdrivers
8-2-13-2
Preparations
Shut down the system and switch off the main breaker.
8-2-13-3
Removal procedure
1.) Remove key caps from the keyboard.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-14
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-8-3
Section 10-5-5
8-2-13-4
Mounting Procedure
Install the new parts in the reverse order of removal.
8 - 46
3
Figure 8-85 Key Caps
Functional Tests
Section
Functional Test / Diagnostic Test
TGC key assy/TGC Knob set, Sub keyboard
encoder Knob Set, keycap set, Main keyboard
Encoder Knob set Function Validation
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
Procedure
Physical Inspection
LOGIQ™ P6/P6 P
S
RO
ERVICE
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-13. Equipment
passed all required tests and is ready for use. "