GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-28
BP128 Assy
8-2-28-1
Tools
•
Common philips screwdrivers
8-2-28-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-28-3
Removal procedure
1.) To replace the BP128 assy, all PWAs should be removed from the nest box.
2.) Open the Side Left Cover and Side Right Cover. Refer to the
page
3.) Open the EMI Cover L and EMI Cover R. Refer to the
4.) Remove all PWAs from the nest box. Refer to the Figure 8-109.
5.) Open the front cover. Refer to the
6.) Remove the P3RLY/PI128II Board. Refer to the
7.) Remove the HDD. Refer to the
8.) Unplug all the cable connectors all in BP128 Assy.
9.) Remove a bracket (1) by looseing 2 screws.Refet to
3
8-47.
Figure 8-109 PWAs in the nest box
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-24 "EMI Cover L and R" on page
8-2-15 "Front Cover" on page
8-2-27 "P3RLY/PI128II Assy" on page
8-2-43 "SATA HDD Assy" on page
Figure 8-110 "BP128 Assy" on page
LOGIQ™ P6/P6 P
S
RO
ERVICE
8-2-14 "Right or Left Side Cover" on
8-49.
8-92.
M
ANUAL
8-62.
8-67.
8-70.