GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-19
Top Bottom Cover
8-2-19-1
Tools
•
Common philips screwdrivers
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-19-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-19-3
Removal procedure
1.) Remove the Dummy Cover L/R. Refer to the
2.) Unscrew 7 screws (1-7) from the bottom of the top cover. Refer to the figure below.
3.) Remove the Top Buttom Cover.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-20
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-8-1
Section 10-5-5
8-2-19-4
Mounting Procedure
Install the new parts in the reverse order of removal.
8 - 54
3
Figure 8-94 Top bottom cover
Functional Tests
Section
Functional Test / Diagnostic Test
Cover Parts Function Validation
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-23 "Dummy Cover L / R" on page
Power On/Boot Up
Power Off / Shutdown
Physical Inspection
LOGIQ™ P6/P6 P
S
RO
ERVICE
8-60.
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-19. Equipment
passed all required tests and is ready for use. "
M
ANUAL