GE LOGIQ P6 Service Manual page 283

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-25
APS II Assy
NOTE:
APS Power Supply can not be adjusted.
8-2-25-1
Tools
Common philips screwdrivers
Nipper
8-2-25-2
Preparations
Shut down the system and switch off the main breaker.
8-2-25-3
Removal procedure
1.) Remove the Side Left Cover. Refer to the
2.) Remove the EMI Cover L. Refer to the
3.) Remove HDD power cable.
4.) Extract 3 power connector from back plane.
5.) Disconnect the AC input connector.
6.) Remove the Side Right Cover. Refer to the
7.) Remove the EMI Cover R. Refer to the
3
AC Input Connector
Figure 8-103 APS IIConnectors of backplane
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P6/P6 P
8-2-14 "Right or Left Side Cover" on page
8-2-24 "EMI Cover L and R" on page
8-2-14 "Right or Left Side Cover" on page
8-2-24 "EMI Cover L and R" on page
S
M
RO
ERVICE
ANUAL
8-47.
8-62.
8-47.
8-62.
8 - 63

Advertisement

Table of Contents
loading

Table of Contents