5-7-5-1
Logs
The seven sub-menus of the Logs category are System, Infomatics, Temperature, Probe, Board, and
DICOM.
NOTE:
Figure 5-26 provides a graphical example of the log entries for the System Logs.
Log table headings for the different logs are as follows:
System
Log entry headings include Time Stamp; Error Level; Package; and Error Message.
Infomatics
Log entry headings include TimeStamp, Revision, PtID, PtDOB, PtSex, PtWeight, PtHeight,
ExamID, Exam Category, ExamCurDate, and ExamStartTime.
Temperature
Log entry headings include Time Stamp; Error Level; Package; L1SYSCON temperature sensor.
Probe
Log entry headings include Time Stamp; Error Level; Package; Error Message; Severity; Revision;
and three (3) new labels that have not yet been named.
DICOM
Log entry headings include Time Stamp; Error Level; Package; and Error Message.
5-7-5-2
Exit
The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page.
5 - 28
3
Figure 5-27 Exit Log Sub-Menu
Section 5-1 - Overview