Aps Ii Assy - GE LOGIQ P6 Service Manual

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
7-8-5

APS II Assy

The APS Assy is located at the No. 5 slot of the NEST Assy.
Table 7-14
LED
DP5 (Top)
LP5
AP6
DP33
SHV+
HVL+
HVH+ (Bottom)
7 - 26
3
LEDs on APS IIAssy
Descriptions
Digital 5V output indicator
Live Power 5V output indicator
Analog 6V output indicator
Digital 3.3V output indicator
SHV+ output indicator status indicator
HVL+ level indicator
HVH+ level indicator
Section 7-8 - LED Descriptions
LOGIQ™ P6/P6 P
S
RO
ERVICE
Normal State
Normally On
Normally On
Normally On
Normally On
Normally On
Normally On in CFM mode or PW mode
Normally On in B mode
M
ANUAL