Diagnostics/Troubleshooting

7-2-3 Collect a Trouble Image with Logs

NOTE:
7-2-3-1 Advanced log options
7-4
If the system should malfunction, press the Alt+D keys
simultaneously. This will collect a screen capture of the image
monitor, system presets and several log files in a date and time
stamped ".zip" file.
This function may also be used to make a Print Screen.
This Alt+D function is available at all times.
When Alt+D is pressed, a menu box appears that allows for:
a place to enter a description od the issue
a checkbox to indicate a System lockup
a chioce to Export a pre-formateed CD-R/DVD-R or save to
the Export directory D: drive (for remote viewing through
InSite)
Figure 7-1. Alt+D Dialog Box
Extensive Log enables the creation of a log file containing
addtional information for the selected functionality.
Options enables creation of a log file based on a selected
bookmark or for a user configurable time frame. Different
type of information can be selected to be part of the log file.
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10