16 Technical data

16.1 Device

Specification
Product class according to Directive
93/42/EEC
Dimensions (W x H x D)
Weight:
Without rechargeable battery
With rechargeable battery
Center of gravity
Operation:
Temperature range
Temperature range for oxygen inlet
tube
Humidity
Air pressure
Altitude above mean sea level
Storage (device)/transport:
Temperature range up to 48 h
Temperature range longer than 48 h
Temperature range for oxygen inlet
tube
Humidity
Air pressure
Altitude above mean sea level
Warm-up time from minimum storage
temperature to standby at 20 °C
Cooling-down time from maximum
storage temperature to standby at
20 °C
Electrical rating
Max. power consumption
Current consumption
Device
IIb
206 mm x 137 mm x 130 mm
1750 g
2100 g
-20 °C to +50 °C
-18 °C to +50 °C
5 % to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
-40 °C to +70 °C
-20 °C to +40 °C (recommended: 0 °C to +25 °C)
-18 °C to +50 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
8 hours
8 hours
Rated voltage: 12 V
Permitted input voltage range: 10 V to 16 V
60 W
0.15 A to 4 A
MEDUVENT Standard
16 Technical data
X = 87 mm
Y = 69 mm
Z = 54 mm
EN
157