Requirements On The User; Using The Device Safely - Weinmann MEDUVENT Standard Instructions For Use Manual

Hide thumbs Also See for MEDUVENT Standard:
Table of Contents

Advertisement

Less risk of hyperventilation
Less risk of undesired hypoventilation
Less risk of gastric insufflation

2.3 Requirements on the user

The user must meet the following requirements:
The user is medically trained and has the necessary technical
knowledge and experience in the emergency medical
treatment of patients.
On the basis of this technical knowledge and experience, the
user is in a position to perform the tasks assigned to him or her
safely, and to independently recognize, assess and avoid
potential risks to him or herself, to the patient or to bystanders.
The user has been trained and has received instruction in the
use of the device.
The user has been trained to apply the necessary hygiene
procedures.

2.4 Using the device safely

Risk of injury from device being used for too long without
further humidification of the respiratory gas!
If the device is used for too long, the patient might be ventilated
with dry gas for too long. This can cause serious or life-threatening
injury to the patient.
 Do not use the device for sustained ventilation (longer than
24 hours).
MEDUVENT Standard
2 Safety
EN
9