PRONOVA SSM 6000 User Manual page 62

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT
Electrical connection values
Electrical connection, voltage suppy
Supply / incoming supply
Power consumption
Fuses / overload protection
Gas connections
Position of gas connections
Gas inlet / safety equipment
Gas and condensate connections, type / design
Measuring-gas processing option
Cooler type
Input dew point / inlet temperature
Outlet dew point
Condensate pump
System monitoring
Characteristic of measuring-gas cooler
800C-07.07.EN
Power cable 3x0.75 mm² with earthed-contact plug
230 VAC 50 Hz / 115VAC 60Hz option
85 VA max.
Miniature fuse 4x20mm / 1 A slow-blow
Underside of control cabinet
F 501 detonation protection
Clamping-ring screw connection PA for hoses d
Peltier cyclone cooler, Duran glass
40°C / 80°C max.
5°C (ex-works setting)
Hose pump, 16ml/min, Novoprene hose
Status displays (LED), alarm at ±3°C
Ausgangstaupunkt in Abhängigkeit vom Eingangstaupunkt und Volumenstrom
10
9
8
7
6
5
4
20
40
60
80
Messgasvolumenstrom in l/h
SSM 6000C User Manual
TECHNICAL SPECIFICATIONS
according to EN 12874
V2A stainless steel
x s = 6 x 1 mm
a
50°C
40°C
30°C
20°C
100
120
140
160