Electrical Connections - PRONOVA SSM 6000 User Manual

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT

2.2.10 Electrical connections

All electrical connections are located on rear side of the analyzer.
Voltage supply
The analyzer system requires up to standard a voltage supply of 230VAC / 50Hz and features a maximum power consumption of
85VA. On option the instrument could be orded wiht initial voltage for 115VAC / 60Hz. A 1.5m long power cable with earthed-contact
plug (3x0,75mm²) serves as a supply cable. The power supply does have a multi-pin with each precision-backup of 4x20 mm 1 A.
The main switch on the front panel is used to turn the system on and off. For further information, please refer to the terminal
diagrams in the appendix.
We recommend installing a mains disconnector switch or a switched receptacle in the power supply circuit near the gas analyzer in
order to be able to disconnect all poles of the gas analyzer from the power supply, if necessary. The mains disconnector switch must
be marked in a manner that clearly shows the equipment to be disconnected.
Signals and interfaces
The SSM 6000 analyzer comes with the following connections for communicating with peripheral equipment.
Plug connector
Description
Analog measured-value outputs 4-20mA, one output per component measured
Port 1
Digital inputs for starting a measuring process and controlling up to four measuring points
12VDC voltage output for controlling the digital inputs
Digital status outputs and limit-value alarms
Port 2
Digital outputs for controlling up to three solenoid valves for the measuring-point switch
12VDC voltage output – for reading/driving the digital outputs
RS 232
RS 232 interface for the output of measured values with time stamp on a PC and/or data logger
Profibus (option)
Interfaces for the output of measuring data and status messages, as well as initialization of a measuring
process and controlling the measuring-point switch via a Profibus system.
Interfaces for the output of measured values and status messages, as well as initialization of a measuring
CAN bus (option)
process and controlling the measuring-point switch via a CAN bus system. This interface is not yet
supported by the software.
For further information, please refer to the terminal diagrams in the appendix and to the "Installation" section.
800C-07.07.EN
SSM 6000C User Manual
HARDWARE DESCRIPTION
21