Checking the interface functions

The analyzer can simulate three analog output values in order to check the analog data output functionality and to adjust this to the
higher-level plant control system, if necessary. This enables adjustment of external isolating switching amplifiers or arithmetic
correction by the controller.
DATA OUTPUT:
_ DATALOGGER
_ ANALOG OUTPUT
_ SENSORDATA
The data output via the RS232 interface is described in more detail in the "Menu and function overview" section.
Further possible function tests:
6.4

Performing a test measurement

The analyzer was calibrated and tested by the manufacturer prior to shipment (refer to test certificate). In order to rule out any risk of
transport damage to the analyzer, one measurement should, if possible, be performed with calibration gas and another
measurement with ambient air when the unit is set into operation. Proceed as follows:
Connect calibration gas with a maximum admission pressure of 0.2 bar to the calibration-gas inlet.
-
Start of a single measurement as described in the following.
-
SETUP MODE:
_ MEASUREMENT
_ CALIBRIATION
_ SYSTEM
SINGLE MEASUREMENT:
PLEASE LEAVE
SETUP
During the measuring process, check and, if necessary, readjust the volume flows at the analyzer. On completion of the measuring
process, the current measured values are displayed and the device returns to stand-by mode.
In order to check the zero points for CH
one measurement with ambient air. For this purpose, disconnect the calibration-gas hose from the measuring-gas input.
If the values measured are within the specified ranges, the system can be used for normal measurements. In the case of larger
deviations, the analyzer should be re-calibrated using a suitable calibration gas as described in the "Calibration" section.
800C-07.07.EN
ANALOG OUTPUT
_ SIMULATION
_ RANGE H2S
_ SETTINGS
Profibus interface
CANBUS interface
RS232 interface
MEAS. OPTIONS:
_ MEAS. INTERVAL
_ ALERT VALUES
_ SINGLE MEAS.
Press the <enter> key in order to confirm the start of the measuring process. When prompted,
operate the key switch in order to exit the setup mode.
, H
S and CO
as well as the sensitivity of the O
4
2
2
SSM 6000C User Manual
SETTING INTO OPERATION
SIMULATION:
SELECTED NO SIM
_
4 mA
_ 20 mA
_ 16 mA
_
0 mA
SINGLE MEASUREMENT:
_ MEASUREPOINT 1
_ START
channel, you should subsequently perform
2
48