SSM 6000
CONT

5 Menu and function overview

The following section describes the functions of all user menus that are available for adjusting the analyzer system to the conditions
of the biogas plant.
The individual user menus can be accessed in setup mode. The setup mode, for its part, is accessed via the key switch.
SETUP MODE:
_ MEASUREMENT
_ CALIBRIATION
_ SYSTEM
5.1

Measuring parameters

The MEASURING PARAMETERS menu contains the following sub-menus.
MEASURING INTERVAL
ALARM VALUES
SINGLE MEASUREMENT
5.1.1
Measuring interval
MEAS.INTERVAL: [min]
_CH4: 720 _O2 : 720
_H2S: 480 _CO2: 720
_RT :12:00 _OFS:
30
With the setting selected above, H
a value of 30 minutes is selected for the offset, the measuring processes are carried out as follows:
1st day
2
day
nd
3
day
rd
Not every interval is permitted. The interval must be a divisor of 1440, i.e. the number of minutes of a day. If the value
entered does not fulfill this requirement, the SSM 6000 automatically uses the nearest setting. However, the measuring
interval should not be set at a value of less than 120 minutes. This corresponds to 12 measurements a day.
If the device is controlled via the digital inputs or via the Profibus interface, an H
order to avoid conflicts with the automatic measuring interval.
800C-07.07.EN
The different menu items can now be selected via the membrane keypad.
For further details concerning operation, please refer to the "Using the gas analyzer" section.
- Setting the measuring interval for every component to be measured.
- Setting the limit-value concentrations for the different gas types.
- Performing one manual measurement.
The measuring interval for H
irrelevant because these other gas components are measured continuously.
In automatic measuring mode, the reference time determines the time when the first measuring
process starts. After this measuring process, the next measuring process is then started xxx minutes
later, depending on the set measuring interval.
The "offset" selection (_OFS: 0000) enables the automatic starting of measuring processes in the
intervals defined for the day with the "minutes" entry edited to this effect.
S are measured every 480 minutes = three times a day at 4:00 a.m., 12:00 noon, and 8:00 p.m. If
2
4:00 / 12:00 / 20:00
4:30 / 12:30 / 20:30
5:00 / 13:00 / 21:00
SSM 6000C User Manual
S is edited in this menu. The settings for the other gas components are
2
S measuring interval of "9999" should be set in
2
MENU AND FUNCTION OVERVIEW