Hardware Installation; Requirements For The Plant Of Installation, Power Supply; Measuring Input And Output Conditions, Calibration Gases - PRONOVA SSM 6000 User Manual

Hide thumbs Also See for SSM 6000:
Table of Contents

Advertisement

SSM 6000
CONT

3 Hardware installation

3.1

Requirements for the plant of installation, power supply

All major system components are installed in a wall-mounted housing with IP20 protection. The device is hence designed for
installation in closed, air-conditioned rooms. The device should be typically installed on a vertical wall in the central machinery room
of the biogas plant as close as possible to the point of sampling.
Air inlet and air outlet openings may not be obstructed by objects or walls in order to ensure sufficient air circulation and to avoid
heat accumulation.
The analyzer system must be protected against adverse ambient conditions, such as:
extreme cold,
-
exposure to heat radiation, for example, sunlight, furnaces, boilers or gas motors,
-
strong temperature fluctuations,
-
accumulation and penetration of dust,
-
aggressive atmosphere and
-
shock/vibration.
-
Climatic conditions
Power supply*
specification is written onto/look at label
*
3.2

Measuring input and output conditions, calibration gases

Measuring input conditions
Instruments with measuring cooler
Instruments without measuring cooler
below
Measuring output conditions
*
Ingress of condensate into the analyzer, for example, in the form of droplets in the biogas pipe/hose, must be avoided under all
circumstances. If necessary, a pre-separator must be installed.
800C-07.07.EN
Air pressure
Relative humidity
Ambient temperature
Input voltage
Power consumption
specification quod vide appliance rating plate
Input dew point of the gas to be measured
Instruments with measuring cooler
Instruments without measuring cooler
Measuring-gas temperature at the input 80°C max.
Pressure at the measuring-gas input **
The measuring gas must be discharged into the outside atmosphere in a non-pressurized
condition using hoses which should be as short as possible. 5m max. / separated
SSM 6000C User Manual
during storage and transport
during operation
230 VAC 50 Hz
115 VAC 60Hz (Option)
85 VA max.
max. 40°C
min. 5°C ambient temperature limit
-200 ... +200 hPa
HARDWARE INSTALLATION
850...1100 hPa
75 % max.
–25...+50 °C
+10...+40 °C
22