Use the <down> and <up> keys in order to set the desired value, and press the <enter> key to confirm your entry. The cursor
advances by one position that can now be changed.
After the last numeral of the input field, the cursor moves to the input field (for example, in front of the text H
<enter> key. The changed value was saved. If you press <esc> in order to cancel the entry before the cursor is positioned in front of
the input field, the changes made so far are not saved.
Not every interval is permitted. The interval must be a divisor of 1440, i.e. the number of minutes of a day. If the value
entered does not fulfill this requirement, the SSM 6000 automatically uses the nearest setting. With the setting selected
above, H
S is measured every 480 minutes and hence three times a day at 4:00 a.m., 12:00 noon, and 8:00 p.m, for
2
example.
Press the <esc> key in order to access the next higher menu.
In the same way, further sub-menus can be accessed by pressing these keys until the desired parameterization level is reached.
Once all settings are made, you can operate the key switch in order to exit the setup mode.
4.2

Starting a measurement

Measurements can be started in different ways.
Manual start of a measuring cycle
In stand-by mode, a measuring cycle can be started at any time by pressing the <enter> key. You are first prompted to select the
measuring point via the <up> and <down> keys. Thereafter, press the <enter> key once again in order to confirm the start of the
measuring process. The unit starts continuous measurements of the CH
interval has expired, H
S measurement starts automatically.
2
Single measurement
Via the menu item MEASURING PARAMETERS -> SINGLE MEASUREMENT, a manual measurement can be performed at any
time while the device is in automatic mode without having to vary the measuring interval or the reference time. Press the <enter> key
in order to confirm the start of the measuring process after you have selected the measuring point. When prompted, operate the key
switch in order to exit the setup mode. The measuring process starts immediately. On completion of the measuring process, the
current measured values are display and the device returns to stand-by mode.
External start via the digital inputs
At plug connector PORT1, the analyzer features four digital inputs which can be used to start an H
time to select the measuring point. The digital optocoupler inputs are assigned as follows and must be activated for around 1.5
seconds:
DI 01
-
measuring point 1
DI 02
-
measuring point 2 / valve 1
DI 03
-
measuring point 3 / valve 2
DI 04
-
measuring point 4 / valve 3
After the start of the measuring process, the analyzer simultaneously activates the corresponding measuring-point valve and/or the
pertinent digital output at plug connector PORT2. The digital inputs are not read again until a H
completed. After an H
S measuring cycle, the continuous measuring mode of the other components continues to be active until
2
another measuring point is selected or until the <esc> key is pressed in order to terminate the measuring cycle.
If the device is controlled via the digital inputs or via the Profibus interface, an H
order to avoid conflicts with the automatic measuring interval.
800C-07.07.EN
, O
and CO
4
2
2
S measuring interval of "9999" should be set in
2
SSM 6000C User Manual
USING THE ANALYZER
S) when you press the
2
components. When the set measuring
S measurement and at the same
2
S measuring process has been
2
32